The cleaning-up project of Panama City and Bay includes: Construction of Health Service networks Construction and rehabilitation of Collectors and Drive lines Interceptor system Plant for wastewater treatment. These elements will be developed to collect, transport and treat wastewater of Panama City which is currently discharged directly into the bay or rivers and streams in […]
FAQ#118 – Is it possible to use the reference reading (zero reading) taken with a digital inclinometer probe as a reference reading for another digital inclinometer probe?
No, this is not possible.Sisgeo calibrates each probe individually, but each probe has its own small different properties, that have a certain influence on the readings. The international standard ISO 18674-3, describing the reference reading, states that:« It is good practice to take an additional reference measurement with an independent probe that can be used as […]
FAQ#111 – Why I cannot read correctly my 4-20mA current loop gauge?
It could be a problem connected to the power supply given by the readout or logger. When the power supply is in the current loop (2-wire gauge) it is necessary to consider the effect of voltage drop across the shunt resistor on the voltage applied to the gauge. For example, suppose a 4-20mA gauge requires […]
FAQ#075 – What parameters are saved in SISGEO digital sensors? What information must the Customer provide when ordering?
All SISGEO digital sensors (e.g. IPI, BH Profile, MD Profile, LT Inclibus, Tilt-meter, RDS, H-Level, etc.) use the Modbus communication protocol over RS-485 serial. SISGEO digital sensors are connected to each other with a single signal cable. The parameters that are configured at the factory for each digital sensor are as follows: Upon confirmation of […]
FAQ#074 – Quel type d’enregistreur de données est nécessaire pour lire les mesures des capteurs numériques SISGEO?
Les relevés des capteurs sont lus en connectant le câble à une unité maîtresse Modbus (à savoir, OMNIAlog ou miniOMNIAlog) avec une interface RS-485 (DCE, Half duplex, no Echo) en mesure de gérer la communication avec le protocole de communication Modbus RTU SISGEO. La documentation relative au protocole de communication de SISGEO est disponible sur […]
Egnatia Odos Road – Athens, Greece
Egnatia Motorway was designed to the specifications of the Trans-European road network. Across Epirus and Northern Greece from Igoumenitsa to Evros, Egnatia Motorway is one of the largest road construction projects in Europe, 670 kilometres long and 24.5 metres wide.The realisation of this motorway requires the construction of: 50 road interchanges; 350 entrance / exit […]
FAQ#073 – Quel est le nombre maximal de capteurs numériques (RS-485, Modbus), le nombre maximal de chaînes et la longueur maximale du câble dans un réseau RS-485?
Last update: October 2021 All the SISGEO digitized sensors utilize a RS485 interface with Modbus protocol. 1.The maximum number of digitized sensors in a RS-485 network are 247 (theoretical) 2.The maximum number of digitized sensors chains that is possible to connect to Modbus master unit (as OMNIAlog or miniOMNIAlog) are 4. 3. About the maximum […]
FAQ#038 – Dois-je mettre en charge le TEL-310S avant de commencer à le configurer et à l’utiliser?
Si le TEL-310S a été livré plus de 2 mois auparavant, il est préférable de le mettre en charge pendant 8 heures avant de commencer à le configurer et à l’utiliser. Cela permettra de recharger la batterie de secours qui peut causer des soucis si elle est complètement déchargée.
FAQ#036 – Quel est le bon mélange pour réaliser le coulis d’un tube inclinomètre?
Il n’existe pas de règle stricte pour chaque type de sol. Cela dépend de nombreux facteurs qui peuvent varier dans chaque situation géologique et géotechnique. Dans tous les cas, le coulis utilisé doit avoir un comportement approchant celui du sol, même s’il n’est pas vraiment possible de faire correspondre leurs caractéristiques parfaitement, en particulier si […]
FAQ#032 – Quelle est la méthode de lecture de la nouvelle génération de postes de lecture et enregistreurs Sisgeo pour les instruments à corde vibrante?
For VW sensors Sisgeo has developed automatic algorithms to search for the resonant frequency in case of environmental noise or poor signal. In this way you can distinguish between the natural frequency of the sensor and the frequency of the noise, thereby reducing measurement errors and false alarms. This method also provides improvement in the […]
FAQ#026 – Que peut-on faire lorsque le tube inclinomètre « flotte » et remonte pendant l’installation?
Si le tube, même rempli d’eau, est plus léger que le coulis de ciment injecté, l’effort d’arrachement du coulis varie avec la hauteur de la colonne de coulis. Si la colonne est courte, l’effort d’arrachement est faible et le tubage peut être maintenu en place par son propre poids avec une très petite force vers […]
FAQ#025 – Comment faire pour obtenir de bons relevés avec une sonde inclinométrique?
Une bonne pratique pour obtenir de bons relevés avec une sonde inclinométrique consiste à suivre ces recommandations simples : – Utilisez la même sonde pour chaque étude, si possible ; – si vous devez utiliser une sonde différente, veillez à noter le numéro de série de la sonde utilisée pour chaque étude, de sorte que […]