FAQ#009 – Can the TEL-310S functioning be influenced from atmospheric conditions?
Any factor (water, mud etc.) that could dirty the internal part of the “SUS” optical sensor unit’s cover can influence the TEL-310S correct functioning. So it is essential to protect properly the “SUS” optical sensor unit from weather conditions, leaving anyway the access possibility for cleaning/maintenance. In any case, TEL-310S has an auto-diagnostics able to […]
FAQ#008 – What is the 0TEL310CAL0 calibration/mounting kit and what is it for?
The 0TEL310CAL0 kit includes a drilling template for the wall where the “ANS” support will be placed, and a template to be placed on the “SUS” optical sensor unit for a possible functional check on field (pendulum’s wire simulation).
FAQ#007 – Are the TEL-310S CUS control unit and the PWS power unit supplied with the fixing kit?
No kit is needed. The “CUS” and “PWS” units are placed in aluminum boxes ready to be fixed on the wall through four M6 screws (not included).
FAQ#006 – Is it possible to use the ANS adjustable support for the pendulum optical manual readout (coordinometer)?
The pendulum readout fixing plate has four holes so it can be fixed, through screws, to the “ANS” adjustable support.
FAQ#005 – Is it possible to fix the ANS adjustable support to the floor?
Since the “ANS” support is articulated, it can be fixed to the floor; this can be useful in case the pendulum’s wire is too far from the walls.
FAQ#004 – Is it possible to place the SUS optical sensor unit with the X axis perpendicular to the wall where the ANS support is fixed?
The TEL-310S telecoordinometer has the following measuring range: To use the “ANS” adjustable support you must have the “X” axis parallel to the wall: so it is not possible to fix the “SUS” optical sensor unit perpendicular to the wall using the “ANS” adjustable support.
FAQ#003 – Which are the needed conditions to use the ANS adjustable support for TEL-310S if you need to monitor the direct pendulum’s wire?
To fix the “SUS” optical sensor unit through the “ANS” support, it is necessary that the direct pendulum’s wire is at a distance of d=230 mm. from the wall. To achieve that, the direct pendulum’s tank has to be placed at a distance of D=20 mm. from the wall: in this way is guaranteed that […]
FAQ#002 – What is the 0TEL310ANS0 adjustable support for TEL-310S and what is it for?
The “ANS” adjustable support (0TEL310ANS0) has been developed to support and fix the “SUS” optical sensor unit (0TEL310SUS0) of the new TEL-310S telecoordinometer. The support is fixed to the wall through two M16 threaded bars, of 370 mm. length (supplied). Once the “SUS” optical sensor unit is fixed to the support, it is possible to […]
FAQ#001 – If the SUS optical sensor unit is removed for maintenance, is it possible to continue the monitoring using the pendulum optical manual readout?
It is possible to use the pendulum optical manual readout instead of TEL-310S “SUS” optical sensor unit without any regulation of the “ANS” support. It is also possible to start the pendulum’s monitoring with the pendulum optical manual readout and then automate the readings with TEL-310S telecoordinometer , without changing its initial zero.
V-NOTCH and Flow Meters

V-NOTCH and Flow Meters Reference products codes: QV, QVML, QVLIV, P252Q Water flow monitoring in open channels is widely employed in environmental and geotechnical field. Leakage measurement is one of the most important indicator of the overall performance of earth/rock-fill and concrete dam. The leakage rate is a function of the water level in the reservoir and […]
Water level meters, Staff gauges

Water level meters, Staff gauges Reference products codes: C112, C112T, HIDR Dipmeters are used for water ground level measurements in standpipes, wells and Casagrande piezometers. It is a battery powered portable device equipped with a dipping probe connected to a graduated cable wrapped on a cable reel. In contact with water the dipping probe activates either an audible […]
Casagrande and standpipe piezometers

Casagrande and standpipe piezometers Reference products codes: P101, P112, P112A, TCH, TFHCasagrande filter unit is used to measure the water pressure in permeable soil. Filter unit is made in syntherised high density polyethylene. It is available in different models to suite all the customer applications. Filter units have threaded cap joint with two 0.5″ twin […]